Weight | 55.11556 lb | ||
accessories | power cable, foot for fixing | ||
electrical system - alarm | refer to alarm function. | ||
electrical system - overcurrent protector | 10 a | ||
facility water system - required flow rate note 5) | stainless steel 303, stainless steel 304 | ||
facility water system - temperature range | 2.1 to 2.6 gpm(8 to 10 l/min) | ||
circulating fluid system - port size | stainless steel 303, stainless steel 304, epdm, ceramics, pps glass 30%, carbon, pe, polyurethane | ||
circulating fluid system - pump capacity | refer to performance chart. | ||
circulating fluid system - heating capacity | 900 w (clear water), 600 w (fluorinert fc-3283) note 2) | ||
circulating fluid system - operating temperature range | 50 to 140f (10 to 60 c) (no condensation) | ||
ambient temperature/humidity | 50 to 95f (10 to 35c), 35 to 80%rh (no condensation) | ||
radiating method | water-cooled | ||
model | hec006-w2b |
safety standards | ce marking | ||
electrical system - communications | rs-232c | ||
electrical system - current consumption | 5 a | ||
electrical system - power supply | single-phase 200 to 220 vac ?10%, 50/60 hz | ||
facility water system - pressure range | in/out: rc3/8 | ||
facility water system - wetted parts material | within 145 psi (1 mpa) | ||
circulating fluid system - tank capacity | approx. 0.8 gal (3 l) | ||
circulating fluid system - temperature stability note 4) | ?0.018 to 0.054f (?0.01 to 0.03c) | ||
circulating fluid system - cooling capacity | 600 w (clear water), 400 w (fluorinert fc-3283) note 2) | ||
circulating fluid system - circulating fluid note 1) | clear water, fluorinated fluid (fluorinert fc-3283, galden? ht135) | ||
control method | cooling/heating automatic shift pid control | ||
cooling method | thermoelectric device (thermo-module) |